Product





 

YSI 2300 STAT Plus™ Glucose & Lactate Analyzer

YSI 2300 STAT Plus™ Glucose & Lactate Analyzer

 

 

 

YSI glucose and lactate analyzers have been recognized standards for the measurement of blood glucose and lactate for more than a quarter century. The YSI 2300 STAT Plus is known for its exceptional speed, performance and ease of use. It is a laboratory glucose and lactate analyzer designed for flexibility. Measure glucose and lactate simultaneously in whole blood or plasma. Alternatively, configure the analyzer to measure glucose only or measure lactate only. Requires only the sensor and reagents for the chemistry you need. The YSI2300 is ideal for:

 

  • -  STAT Lab or Central Lab
  • -  Diabetes Evaluation
  • -  Diabetes / Endocrinology Research
  • -  Handheld Monitor Development
  • -  Human Performance Evaluation
  • -  Exercise / Athlete Evaluation
  • -  Cardiac / Pulmonary Rehabilitation
  • -  CNeonatology / Pediatrics
  • -  General Metabolic Research
  • -  Veterinary Medicine & Research
  • -  Pharmaceutical Quality Assurance

 

 

 

Product Specifications
 
 

Standard Features:

In addition to speed, autocalibration,and small sample size,
the YSI 2300 STAT PLUS features:

• Lower cost per test. No need to renew enzymes after each
test, minimizing the use of costly reagents and supplies

• Whole blood capability. The only dual channel glucose and
lactate analyzer that analyzes whole blood----no need to spin
blood.

• Simple, safe operation.Menu-driven software guides
operators through each step.

• Batch runs. 24-position turntable provides batch operation
capability for plasma or serum. Each batch and sample
is assigned an identification number with date and time.

• Interference rejection.
Eliminates most of the interferences that plague other methods.

• Broad measurement range.
Eliminates the need for sampledilution.

 

Option & Accessories

YSI 2300 STAT Plus Glucose & Lactate Analyzer

Product Specifications
Sample size:

http://biotech.ysi.com/icons/ecblank.gif
25 µL (aspirated volume)

Response time
(from test tube holder)
Normal mode:

http://biotech.ysi.com/icons/ecblank.gif
Displayed and printed result in 65 seconds.
Sample to sample interval is 100 seconds.

Screen mode: Displayed and printed result in 45 seconds.
Sample to sample interval is 70 seconds.
Measurement range:
Normal mode:
Glucose: to 900 mg/dL (9000 mg/L, 50.0 mmol/L)
Lactate: to 30.0 mmol/L (267 mg/dL, 2670 mg/L)
Screen mode: Glucose: to 500 mg/dL (5000 mg/L, 27.8 mmol/L)
Lactate: to 15.0 mmol/L (134 mg/dL, 1335 mg/L)
Precision:
Glucose:
Whichever is larger, ±2% of reading or 2.5 mg/dL
(25 mg/L or 0.2 mmol/L)
Lactate: Whichever is larger, ±2% of reading or 0.1 mmol/L
(1 mg/dL or 10 mg/L)
Calibration point:
Glucose:
180 mg/dL (1800 mg/L, 10.0 mmol/L)
Lactate: 5.00 mmol/L (45 mg/dL, 445 mg/L)
Membrane working life (typical)
Glucose:
21 days
Lactate: 14 days
Operating temperature: 15 to 35°C
Humidity: 10 to 90% noncondensing
Power: 110-120 VAC or 220-240 VAC, 50-60 Hz
Size: 25.4 x 35.6 x 35.6 cm, 11.4 kg
10 x 14 x 14 inches, 25 lbs
Regulatory Compliance: CE, FDA Registered Class II (862.1345), CSA
CE Device Category: IVD
YSI 2710 Turntable
Sample containers: 10 to 16 mm dia, 20 to 100 mm height
Sample wheel: 19.1 dia x 1.9 cm
(7.5 dia x 0.75 in)
Size: 23.5 x 20.3 x 12.7 cm, 2.5 kg
(9.25 x 8 x 5 in, 5.5 lbs)

 

Application Notes

Options & Accessories

Item

Part Number

STAT Plus Glucose and Lactate Analyzer

2300D

24 Position Turntable

2710

IQ/OQ Documentation Package

2399

Preservative Kit

2315

Cell Lysing (8 packets/4 liters)

1515

Membrane Installation Solution

2392

Membrane Check Solution

2363

Printer Paper 5-pack

2751

Preventative Maintenance Kit 2788
Data Logging Software 2340

Related Reagents

 

Chemistry Membrane System Buffer Calibration Standard Linearity Check Solution Starter Kit
Glucose 2365
(4-pack)
2357
(dry mix) makes 4 liters
2747 1531 (125 ml) 9.00 g/L or 50.0 mmole/L 2324
L-Lactate 2329
(4-pack)
2357
(dry mix) makes 4 liters
2747 1530 (125 ml) 2.70 g/L or 30.0 mmole/L 2325
Glucose
Lactate

2365
2329
(4-pack)

2357
(dry mix) makes 4 liters
2747 1530
1531
2323

IQ/OQ Documentation for your c-GMP Needs

YSI Life Sciences provides documentation that helps you with installation and operational qualifications if you are incorporating the YSI 2300 into manufacturing or laboratory processes that require FDA regulation. Both hard copy and digital versions are available to incorporate into your documentation. If you need more than documents ask YSI about on-site services for IQ/OQ implementation.

 

Catalog & References

 

http://www.ysilifesciences.com/index.php?page=clinical-research-and-diagnostics